Patient Timeline
A patient enters the study after SDH evacuation with a ClearFit implant and is followed through their normal imaging schedule for roughly 3–12 months. Each clinically ordered CT or MRI is paired with a research TCUS; nothing in the clinical schedule changes.
- 1
Surgery (standard care).
SDH evacuation by burr hole, craniotomy, or craniectomy, with ClearFit reconstruction. Window size, location, and implant model are recorded from operative and device records.
- 2
Screening.
Site PI confirms: age ≥ 18; SDH as the index indication; usable sonic window ≥ 2.0 cm; planned postoperative CT/MRI; no active scalp or wound infection over the scanning site; not pregnant. Poor visibility at a later exam is an evaluability outcome, not a reason to remove the patient.
- 3
Consent.
Written consent from the patient or legally authorized representative. Surrogate consent early postoperatively if needed; patient re-consents when capacity returns.
- 4
Enrollment.
Study ID assigned. Baseline data entered. Coordinator begins tracking planned imaging.
- 5
First paired exam.
Pairing rules applyTCUS as close as feasible to the first postoperative CT/MRI, preferably same day. Operator maps the window and acquires the exam without reviewing the CT/MRI.
- 6
Upload.
Coordinator de-identifies and uploads TCUS, CT/MRI DICOM, and the official report within 2 working days.
- 7
Central reads.
TCUS reader (blinded to CT/MRI) and neuroradiology reader (blinded to TCUS) each lock within 5 working days of a complete upload.
- 8
Serial paired exams.
Pairing rules applyRepeated at each clinical follow-up imaging point (typically ~1, 3, and 6 months, or as clinically indicated) and at any urgent imaging. Same landmark and acquisition settings as the first evaluable exam.
- 9
Follow-up and safety.
Clinical outcomes, return to surgery, and adverse events tracked through the clinical course.
- 10
Data lock and analysis.
After both reads lock for every pair, the core analyst registers the TCUS field to CT/MRI, clips the segmentation to the matched region, and the coordinating center compares impressions. Adjudication only where triggered.